Wed, 19 Feb 1997 14:06:48 +0000

Dear FlOWers,

We are running a BD-Vantage equiped with a Quadra and Cellquest
software. We too had connection problems, mainly after running the
cytometer in sorting mode for > 1 hour. The error message was : 'GPIO error'.
Shutting off both cytometer and computer was the only thing that
worked. The BD techniqual service has tried everything, including
changing the computer and interface boards, with practically no
results. The only trick we found, was to open the cytometer's door
during sorting (so that there is no overheating of the electronics??) This has
significantly improved our problems...

Bart Lambrecht
Department of Respiratory Diseases
University Ghent
Belgium
bart.lambrcht@rug.ac.be


Home Page Table of Contents Sponsors E-Mail Archive Web Sites

CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone: (765)-494-0757; FAX(765) 494-0517; Web http://www.cyto.purdue.edu , EMAIL cdrom3@flowcyt.cyto.purdue.edu