clean machine

paula fukushima (pif@box-p.nih.gov)
Fri, 14 Feb 1997 14:51:23 -0500

Dear Flow Users,
We use 2% Contrad(Curtin-Matheson)in distilled water
(5 minutes) followed by distilled water(10 minutes) for cleaning lines and
flow cell. It works great with major clogs dissolving in 5 minutes. It is
cheap ~$20.00 per gallon and does not corrode parts. Do not use greater than
2% or it will be hard to get rid of the bubbles.
Paula

Paula Fukushima
NIH, NCI, LAB OF PATHOLOGY
BETHESDA, MD 20892


Home Page Table of Contents Sponsors E-Mail Archive Web Sites

CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone: (765)-494-0757; FAX(765) 494-0517; Web http://www.cyto.purdue.edu , EMAIL cdrom3@flowcyt.cyto.purdue.edu