Dear Flow Users,
We use 2% Contrad(Curtin-Matheson)in distilled water
(5 minutes) followed by distilled water(10 minutes) for cleaning lines and
flow cell. It works great with major clogs dissolving in 5 minutes. It is
cheap ~$20.00 per gallon and does not corrode parts. Do not use greater than
2% or it will be hard to get rid of the bubbles.
Paula
Paula Fukushima
NIH, NCI, LAB OF PATHOLOGY
BETHESDA, MD 20892